The unit eliminates powder contamination to protect the operator and the surrounding environment. It is designed to handle powder under aseptic conditions for optional G.M.P.
• No Cross Contamination
• No Risk of inhalation of powders
• Easy to Clean
• Easy HEPA Filter exchange from underneath unit
Technical Specifications
• Air cleanliness : ISO Class 5 / Class 100 / M 3.5 AREA
• Air Velocities : 0.45 ± 0.05 mps (90 ± 20 FPM)
• Air Flow: Vertical
• Noise Level: Less than 65 dB
• Vibration Level: Minimum
• POWER SUPPLY: 230V AC, 1-Ø, 50HZ
• MOC : GI Powder Coated / SS-304 / SS-316
• Inside: Side panel with all three side with coving
• Three Stage filtration
• EU4 – Pre-filters (90% down to 10 microns)
• EU7 - Intermediate Filter (95% down to 3 microns)
• EU14 - HEPA Filter - Supply / Exhaust (99.999% down to 0.3 microns)
• Statically & Dynamically balanced Motor-blower with isolator to reduce noise level
• Fluorescent Lights with Milky White Diffuser
• Differential Pressure Gauges: 02 Nos. (Intermediate & HEPA Filter)
• ON/OFF Switches (BL / FL / UV)
• Auxiliary Powder Socket: 5/15 Amp for external equipment
• DOP Port
• DQ, IQ, OQ, PQ Documentation
Options and Accessories
• A wide range of configurations to suit any containment application:
• Double skinned PUF Insulated side panels with/without view windows
• Stainless steel perforated grille for HEPA Filter
• Front Flexible Strip PVC Curtains
• SS Perforated Platform for keeping Weighing Scale (Fix /Moveable)
• Velocity display with High / Low Alarms
• Filter chocking alarms
• FLP Electrical Goods for Explosion