Usage/Application | Industrial |
Operating Range | +- 6 V |
Warranty | 3 Year |
Power Supply | 220 VAC |
BT3554, BT3554-10 | |
Resistance measurement range | 3 mΩ (max. display 3.100 mΩ, resolution 1 μΩ) to 3 Ω (max. display 3.100 Ω, resolution 1 mΩ), 4 ranges Accuracy: ±0.8 % rdg. ±6 dgt. (3 mΩ range: ±1.0 % rdg. ±8 dgt.) Testing source frequency: 1 kHz ±30 Hz, With function for avoiding noise frequency enabled: 1 kHz ±80 Hz, Testing current: 160 mA (3m/30 mΩ range), 16 mA (300 mΩ range), 1.6 mA (3 Ω range), Open terminal Voltage: 5 V peak |
Voltage measurement range | ± 6 V (max. display ±6.000 V, resolution: 1 mV) to ± 60 V (max. display ±60.00 V, resolution: 10 mV), 2 ranges, Accuracy: ±0.08 % rdg. ±6 dgt. |
Absolute max. input voltage | 60 V DC max. (No AC input) |
Display update rate | Approx. 3 times/s |
Comparator functions | Resistance warning limit / failure limit, and voltage warning limit setting; Number of comparator setting: 200 sets, Beeper sound, PASS/WARNING or FAIL |
Data storage | Max. storable data: 6000 sets. (Saved items: Date, time, resistance value, voltage value, temperature, comparator setting values, and comparator judgement.) |
Communication interface | Via USB (forward the data to the PC, bundled software) No Bluetooth® Wireless technology |
Other functions | Temperature measurement (-10.0 to 60.0 °C), Zero-adjustment, Hold, Auto-hold, Auto-memory, Auto-power-save, Clock |
Power supply | LR6 (AA) Alkaline dry battery ×8, Continuous operating time: 8.5 hr |
Dimensions and mass | 199 mm (7.83 in)W × 132 mm (5.20 in)H × 60.6 mm (2.39 in)D (with protector), 937 g (33.1 oz) (including batteries and protector) |
Accessories | Zero adjustment board ×1, Instruction manual ×1, Application software CD ×1, Power-on option sticker ×1, Carrying case ×1, LR6 (AA) alkaline batteries ×8, Spare fuse ×1, USB cable ×1, Neck strap ×1, Protector ×1 |
Pin Type Lead 9465-10 ×1 (Bundled with the BT3554), Pin Type Lead L2020 ×1 (Bundled with the BT3554-10) |
Usage/Application | Industrial |
Operating Range | +- 6 V |
Warranty | 3 Year |
Power Supply | 220 VAC |
BT3554, BT3554-10 | |
Resistance measurement range | 3 mΩ (max. display 3.100 mΩ, resolution 1 μΩ) to 3 Ω (max. display 3.100 Ω, resolution 1 mΩ), 4 ranges Accuracy: ±0.8 % rdg. ±6 dgt. (3 mΩ range: ±1.0 % rdg. ±8 dgt.) Testing source frequency: 1 kHz ±30 Hz, With function for avoiding noise frequency enabled: 1 kHz ±80 Hz, Testing current: 160 mA (3m/30 mΩ range), 16 mA (300 mΩ range), 1.6 mA (3 Ω range), Open terminal Voltage: 5 V peak |
Voltage measurement range | ± 6 V (max. display ±6.000 V, resolution: 1 mV) to ± 60 V (max. display ±60.00 V, resolution: 10 mV), 2 ranges, Accuracy: ±0.08 % rdg. ±6 dgt. |
Absolute max. input voltage | 60 V DC max. (No AC input) |
Display update rate | Approx. 3 times/s |
Comparator functions | Resistance warning limit / failure limit, and voltage warning limit setting; Number of comparator setting: 200 sets, Beeper sound, PASS/WARNING or FAIL |
Data storage | Max. storable data: 6000 sets. (Saved items: Date, time, resistance value, voltage value, temperature, comparator setting values, and comparator judgement.) |
Communication interface | Via USB (forward the data to the PC, bundled software) No Bluetooth® Wireless technology |
Other functions | Temperature measurement (-10.0 to 60.0 °C), Zero-adjustment, Hold, Auto-hold, Auto-memory, Auto-power-save, Clock |
Power supply | LR6 (AA) Alkaline dry battery ×8, Continuous operating time: 8.5 hr |
Dimensions and mass | 199 mm (7.83 in)W × 132 mm (5.20 in)H × 60.6 mm (2.39 in)D (with protector), 937 g (33.1 oz) (including batteries and protector) |
Accessories | Zero adjustment board ×1, Instruction manual ×1, Application software CD ×1, Power-on option sticker ×1, Carrying case ×1, LR6 (AA) alkaline batteries ×8, Spare fuse ×1, USB cable ×1, Neck strap ×1, Protector ×1 |
Pin Type Lead 9465-10 ×1 (Bundled with the BT3554), Pin Type Lead L2020 ×1 (Bundled with the BT3554-10) |