Description

The SERVO-i Ventilator family addresses the very different requirements of neonatal, pediatric and adult needs from a single Drager EVITA 4 Ventilator platform. All four configurations are the same ventilator, equipped with different functions. They can be customized and upgraded with different options for future needs. (SERVO-i Adult and SERVO-i Infant can also be upgraded to SERVO-i Universal.)

Minimum Order Quantity: 1
1 Pack contains : 1 Unit

Parameter settings: Setting range:

Parameter:
Inspiratory tidal volume (ml): 100 – 2000
Inspiratory minute volume 0.5 – 60
(l/min):
Apnea, time to alarm (s): 15 – 45
Automode Trigger timeout (s): 7 – 12
PC/PS above PEEP (cmH2O) 0 – (120 - PEEP)
PC/PS above PEEP in NIV 0 – (32 - PEEP)
(cmH2O)
PEEP (cmH2O) 0 – 50
PEEP in NIV (cmH2O) 2 – 20
CPAP pressure (cmH2O): –
CMV frequency (breaths/min): 4 – 100
SIMV frequency (breaths/min): 1 – 60
Breath cycle time, SIMV (s): 1 – 15
PHigh (cmH2O) (PEEP +1) – 50
THigh (s): 0.2 – 10
TPEEP (s): 0.2 – 10
PS above PHigh (cmH2O): 0 – (120 - PHigh)
O2 concentration (%): 21 – 100
I:E ratio: 1:10 – 4:1

Parameter: Setting range:

Oxygen breaths: 100% for 1 minute
Initiation of 1 breath (In SIMV
mode initiation of 1 mandatory
breath) Start breath:
Pause hold: Insp. or exp (0 – 30 seconds)
Alarm silence/reset: 2 minute silence and reset of
latched alarms
Compliance compensation: On/Off
Automode (optional with Automode On/Off
Universal Basic Edition):
Nebulizer (optional) 5 - 30 min./Continuous/Off
• Backup ventilation Backup On/Off

Backup settings Parameter: Setting range:

Inspiratory tidal volume (ml): 100 - 2000
PC above PEEP (cmH2O) 5 – (120 - PEEP)
PC above PEEP in NIV 5 – (32 - PEEP)
CMV frequency (breaths/min): 4 – 100
I:E ratio: 1:10 – 4:1
TInsp (s): 0.1 – 5

SELLER: Meditech Electronics

Drager EVITA 4 - Ventilator

Product code (MPN): ITB-BC10ED8B
Availability:In Stock
₹ 410000.00  Incl. GST ₹ 347457.63 Excl. GST
  • Description
    The SERVO-i Ventilator family addresses the very different requirements of neonatal, pediatric and adult needs from a single Drager EVITA 4 Ventilator platform. All four configurations are the same ventilator, equipped with different functions. They can be customized and upgraded with different options for future needs. (SERVO-i Adult and SERVO-i Infant can also be upgraded to SERVO-i Universal.)

    Minimum Order Quantity: 1
    1 Pack contains : 1 Unit

    Parameter settings: Setting range:

    Parameter:
    Inspiratory tidal volume (ml): 100 – 2000
    Inspiratory minute volume 0.5 – 60
    (l/min):
    Apnea, time to alarm (s): 15 – 45
    Automode Trigger timeout (s): 7 – 12
    PC/PS above PEEP (cmH2O) 0 – (120 - PEEP)
    PC/PS above PEEP in NIV 0 – (32 - PEEP)
    (cmH2O)
    PEEP (cmH2O) 0 – 50
    PEEP in NIV (cmH2O) 2 – 20
    CPAP pressure (cmH2O): –
    CMV frequency (breaths/min): 4 – 100
    SIMV frequency (breaths/min): 1 – 60
    Breath cycle time, SIMV (s): 1 – 15
    PHigh (cmH2O) (PEEP +1) – 50
    THigh (s): 0.2 – 10
    TPEEP (s): 0.2 – 10
    PS above PHigh (cmH2O): 0 – (120 - PHigh)
    O2 concentration (%): 21 – 100
    I:E ratio: 1:10 – 4:1

    Parameter: Setting range:

    Oxygen breaths: 100% for 1 minute
    Initiation of 1 breath (In SIMV
    mode initiation of 1 mandatory
    breath) Start breath:
    Pause hold: Insp. or exp (0 – 30 seconds)
    Alarm silence/reset: 2 minute silence and reset of
    latched alarms
    Compliance compensation: On/Off
    Automode (optional with Automode On/Off
    Universal Basic Edition):
    Nebulizer (optional) 5 - 30 min./Continuous/Off
    • Backup ventilation Backup On/Off

    Backup settings Parameter: Setting range:

    Inspiratory tidal volume (ml): 100 - 2000
    PC above PEEP (cmH2O) 5 – (120 - PEEP)
    PC above PEEP in NIV 5 – (32 - PEEP)
    CMV frequency (breaths/min): 4 – 100
    I:E ratio: 1:10 – 4:1
    TInsp (s): 0.1 – 5